Abstract¶
Purpose: T1 mapping is a widely used quantitative MRI technique, but its tissue-specific values remain inconsistent across protocols, sites, and vendors. The ISMRM Reproducible Research and Quantitative MR study groups jointly launched a challenge to assess the reproducibility of a well-established inversion recovery T1 mapping technique, with acquisition details published solely as a PDF, on a standardized phantom and in human brains.
Methods: The challenge used the acquisition protocol from Barral et al. 2010. Researchers collected T1 mapping data on the ISMRM/NIST phantom and/or in human brains. Data submission, pipeline development, and analysis were conducted using open-source platforms. Inter-submission and intra-submission comparisons were performed.
Results: Eighteen submissions (39 phantom and 56 human datasets) on scanners by three MRI vendors were collected at 3T (except one, at 0.35T). The mean coefficient of variation (COV) was 6.1% for inter-submission phantom measurements, and 2.9% for intra-submission measurements. For humans, inter-/intra-submission COV was 5.9/3.2% in the genu and 16/6.9% in the cortex. An interactive dashboard for data visualization was also developed: https://
Conclusion: The T1 inter-submission variability was twice as high as the intra-submission variability in both phantoms and human brains, indicating that the acquisition details in the selected paper were insufficient to reproduce a quantitative MRI protocol. This study reports the inherent uncertainty in T1 measures across independent research groups, bringing us one step closer to a practical clinical baseline of T1 variations in vivo.
Dashboard: Challenge Submissions
+++
1| introduction¶
Significant challenges exist in the reproducibility of quantitative MRI (qMRI) Keenan et al., 2019. Despite its promise of improving the specificity and reproducibility of MRI acquisitions, few qMRI techniques have been integrated into clinical practice. Even the most fundamental MR parameters cannot be measured with sufficient reproducibility and precision across clinical scanners to pass the second of six stages of technical assessment for clinical biomarkers Fryback & Thornbury, 1991Schweitzer, 2016Seiberlich et al., 2020. Half a century has passed since the first quantitative T1 (spin-lattice relaxation time) measurements were first reported as a potential biomarker for tumors Damadian, 1971, followed shortly thereafter by the first in vivo T1 maps Pykett & Mansfield, 1978 of tumors, but there is still disagreement in reported values for this fundamental parameter across different sites, vendors, and measurement techniques Stikov et al., 2015.
Among fundamental MRI parameters, T1 holds significant importance Boudreau et al., 2020. T1 represents the time constant for recovery of equilibrium longitudinal magnetization. T1 values will vary depending on the molecular mobility and magnetic field strength Bottomley et al., 1984Wansapura et al., 1999Dieringer et al., 2014. Knowledge of the T1 values for tissue is crucial for optimizing clinical MRI sequences for contrast and time efficiency Ernst & Anderson, 1966Redpath & Smith, 1994Tofts, 1997 and to calibrate other quantitative MRI techniques Sled & Pike, 2001Yuan et al., 2012. Inversion recovery (IR) Drain, 1949Hahn, 1949 is considered the gold standard for T1 measurement due to its robustness against effects like B1 inhomogeneity Stikov et al., 2015, but its long acquisition times limit the clinical use of IR for T1 mapping Stikov et al., 2015. In practice, IR is often used as a reference for validating other T1 mapping techniques, such as variable flip angle imaging (VFA) Fram et al., 1987Deoni et al., 2003Cheng & Wright, 2006, Look-Locker Look & Locker, 1970Messroghli et al., 2004Piechnik et al., 2010, and MP2RAGE Marques et al., 2010Marques & Gruetter, 2013.
In ongoing efforts to standardize T1 mapping methods, researchers have been actively developing quantitative MRI phantoms Keenan et al., 2018. The International Society for Magnetic Resonance in Medicine (ISMRM) and the National Institute of Standards and Technology (NIST) collaborated on a standard system phantom Stupic et al., 2021, which was subsequently commercialized (Premium System Phantom, CaliberMRI, Boulder, Colorado). This phantom has since been used in large multicenter studies, such as Bane et al. Bane et al., 2018 which concluded that acquisition protocols and field strength influence accuracy, repeatability, and interplatform reproducibility. Another NIST-led study Keenan et al., 2021 found no significant T1 discrepancies among measurements using NIST protocols across 27 MRI systems from three vendors at two clinical field strengths.
The 2020 ISMRM reproducibility challenge [1] posed a slightly different question: can an imaging protocol, independently implemented at multiple centers, consistently measure one of the fundamental MRI parameters (T1)? To assess this, we proposed using inversion recovery on a standardized phantom (ISMRM/NIST system phantom) and the healthy human brain. Specifically, this challenge explored whether the acquisition details provided in a seminal paper on T1 mapping Barral et al., 2010 is sufficient to ensure the reproducibility across independent research groups.
2| methods¶
2.1| Phantom and human data¶
The challenge asked researchers with access to the ISMRM/NIST system phantom Stupic et al., 2021 (Premium System Phantom, CaliberMRI, Boulder, Colorado) to measure T1 maps of the phantom’s T1 plate (Table 1). Researchers who participated in the challenge were instructed to record the temperature before and after scanning the phantom using the phantom’s internal thermometer. Instructions for positioning and setting up the phantom were devised by NIST and were provided to researchers through the NIST website [2]. In brief, the instructions explained how to orient the phantom and how long the phantom should be in the scanner room prior to scanning to achieve thermal equilibrium.
Table 1:Reference [3] T1 values of the NiCl2 array of the standard system phantom (for both phantom versions) measured at 20 °C and 3T. Phantoms with serial numbers 0042 or less are referred to as “Version 1”, and those 0043 or greater are “Version 2”.
Sphere | Version 1 (ms) | Version 2 (ms) |
|---|---|---|
1 | 1989 ± 1.0 | 1883.97 ± 30.32 |
2 | 1454 ± 2.5 | 1330.16 ± 20.41 |
3 | 984.1 ± 0.33 | 987.27 ± 14.22 |
4 | 706 ± 1.0 | 690.08 ± 10.12 |
5 | 496.7 ± 0.41 | 484.97 ± 7.06 |
6 | 351.5 ± 0.91 | 341.58 ± 4.97 |
7 | 247.13 ± 0.086 | 240.86 ± 3.51 |
8 | 175.3 ± 0.11 | 174.95 ± 2.48 |
9 | 125.9 ± 0.33 | 121.08 ± 1.75 |
10 | 89.0 ± 0.17 | 85.75 ± 1.24 |
11 | 62.7 ± 0.13 | 60.21 ± 0.87 |
12 | 44.53 ± 0.090 | 42.89 ± 0.44 |
13 | 30.84 ± 0.016 | 30.40 ± 0.62 |
14 | 21.719 ± 0.005 | 21.44 ± 0.31 |
Researchers were also instructed to collect T1 maps in healthy human brains, and were asked to measure a single slice positioned parallel to the anterior commissure - posterior commissure (AC-PC) line. Prior to imaging, the imaging subjects consented [4] to share their de-identified data with the challenge organizers and on the Open Science Framework (OSF.io) website. As the submitted data was a single slice, the researchers were not instructed to de-face the data of their imaging subjects. Researchers submitting human data provided written confirmation to the organizers that their data was acquired in accordance with their institutional ethics committee (or equivalent regulatory body) and that the subjects had consented to data sharing as outlined in the challenge.
2.2| MRI Acquisition Protocol¶
Researchers followed the inversion recovery T1 mapping protocol optimized for the human brain as described in the paper published by Barral et al. Barral et al., 2010, which used: TR = 2550 ms, TIs = 50, 400, 1100, 2500 ms, TE = 14 ms, 2 mm slice thickness and 1×1 mm2 in-plane resolution. Note that this protocol is not suitable for fitting models that assume TR > 5T1. Instead, the more general Barral et al. Barral et al., 2010 fitting model described in Section 2.4 can be used, and this model is compatible with both magnitude-only and complex data. Researchers were instructed to closely adhere to this protocol and report any deviations due to technical limitations.
2.3| Data Submissions¶
Data submissions for the challenge were handled through a GitHub repository (https://

Figure 1. List of the datasets submitted to the challenge. Submissions that included phantom data are shown in a), and those that included human brain data are shown in b). For the phantom (panel a), each submission acquired its data using a single phantom, but some researchers shared the same physical phantom with each other. Green indicates submissions used for inter-submission analyses, and orange indicates the sites used for intra-submission analyses. T1 maps used in the calculations of inter- (green) and intra- (orange) submission coefficients of variation (COV) are indicated with asterisks. Images c) and d) illustrate the ROI choice in phantoms and humans.
2.4| Fitting Model and Pipeline¶
A reduced-dimension non-linear least squares (RD-NLS) approach was used to fit the complex general inversion recovery signal equation:
where a and b are complex constants. This approach, developed by Barral et al. Barral et al., 2010, offers a model for the general T1 signal equation without relying on the long-TR approximation. The a and b constants inherently factor TR in them, as well as other imaging parameters such as excitation pulse angle, inversion pulse flip angles, TR, TE, TI, and a constant that has contributions from T2 and the receive coil sensitivity. Barral et al. [31] shared their MATLAB (MathWorks, Natick, MA) code for the fitting algorithm used in their paper [7]. Magnitude-only data were fitted to a modified version of Eq. 1 (Eq. 15 of Barral et al. 2010) with signal-polarity restoration by finding the signal minima, fitting the inversion recovery curve for two cases (data points for TI < TIminimum flipped, and data points for TI ≤ TIminimum flipped), and selecting the case that resulted in the best fit based on minimizing the residual between the model and the measurements [8]. This code is available as part of the open-source software qMRLab Karakuzu et al., 2020Cabana et al., 2015, which provides a standardized application program interface (API) to call the fitting in MATLAB/Octave scripts.
A data processing pipeline was written using MATLAB/Octave in a Jupyter Notebook. This pipeline downloads every dataset from OSF.io (osf.io/ywc9g/), loads its configuration file, fits the T1 maps, and then saves them to NIfTI and PNG formats. The code is available on GitHub (https://
2.5| Image Labeling & Registration¶
The T1 plate (NiCl2 array) of the phantom has 14 spheres that were labeled as the regions-of-interest (ROI) using a numerical mask template created in MATLAB, provided by NIST researchers (Figure 1-c). To avoid potential edge effects in the T1 maps, the ROI labels were reduced to 60% of the expected sphere diameter. A registration pipeline in Python using the Advanced Normalization Tools (ANTs) Avants et al. (2009) was developed and shared in the analysis repository of our GitHub organization (https://
For human data, manual ROIs were segmented by a single researcher (M.B., 11+ years of neuroimaging experience) using FSLeyes McCarthy (2019) in four regions (Figure 1-d): located in the genu, splenium, deep gray matter, and cortical gray matter. Automatic segmentation was not used because the data were single-slice and there was inconsistent slice positioning between datasets.
2.5.1| Analysis and Statistics¶
Analysis code and scripts were developed and shared in a version-controlled public GitHub repository [9]. The T1 fitting and data analysis were performed by M.B., one of the challenge organizers. Computational environment requirements were containerized in Docker Merkel, 2014Boettiger, 2015 to create an executable environment that allows for analysis reproduction in a web browser via MyBinder [10] Project Jupyter et al., 2018. Backend Python files handled reference data, database operations, ROI masking, and general analysis tools. Configuration files handled dataset information, and the datasets were downloaded and pooled using a script (make_pooled_datasets.py). The databases were created using a reproducible Jupyter Notebook script and subsequently saved in the repository.
The mean T1 values of the ISMRM/NIST phantom data for each ROI were compared with temperature-corrected reference values and visualized in three different types of plots (linear axes, log-log axes, and error relative to the reference value). Temperature correction involved nonlinear interpolation [11] of a NIST reference table of T1 values for temperatures ranging from 16 °C to 26 °C (2 °C intervals) as specified in the phantom’s technical specifications. For the human datasets, the mean and standard deviations for each tissue ROI were calculated from all submissions across all sites. Two submissions (one of phantom data – submission 6 in Figure 1-a, and one of human data – submission 18 in Figure 1-b) were received that measured large T1 mapping datasets. Submission 6 consisted of data from one traveling phantom acquired at seven Philips imaging sites, and submission 18 was a large cohort of volunteers that were imaged on two 3T scanners, one GE and one Philips. These datasets (identified in orange in Figures 1, 3, and 4) were used to calculate intra-submission coefficients of variation (COV) (one per scanner/volunteer, identified by asterisks in Figure 1-a and 1-b), and inter-submission COVs were calculated using one T1 map from each of these (orange) along with one from all other submissions [12] (identified as green in Figures 1, 3, and 4, and the T1 maps used in those COV calculations are also indicated with asterisks in Figure 1-a and 1-b). All quality assurance and analysis plot images were stored in the repository. Additionally, the database files of ROI values and acquisition details for all submissions were also stored in the repository.
2.5.2| Dashboard¶
To widely disseminate the challenge results, a web-based dashboard was developed (Figure 2, https://
Figure 2 Dashboard. a) Welcome page listing all the sites, the types of subject, and scanner vendor, and the relationship between the three. b) The phantom tab for a selected ROI, and c) The in vivo tab for a selected ROI. Link: [https://rrsg2020.dashboards.neurolibre.org](https://rrsg2020.dashboards.neurolibre.org)
3| results¶
Figure 3 presents a comprehensive overview of the challenge results through violin plots, depicting inter- and intra- submission comparisons in both phantoms (a) and human (b) datasets. For the phantom (Figure 3-a), the average inter-submission COV for the first five spheres, representing the expected T1 value range in the human brain (approximately 500 to 2000 ms) was 6.1%. By addressing outliers from two sites associated with specific challenges for sphere 4 (signal null near a TI), the mean inter-submission COV was reduced to 4.1%. One participant (submission 6, Figure 1) measured T1 maps using a consistent protocol at 7 different sites, and the mean intra-submission COV across the first five spheres for this submission was calculated to be 2.9%.
For the human datasets (Figure 3-b), inter-submission COVs for independently-implemented imaging protocols were 5.9% for genu, 10.6 % for splenium, 16 % for cortical GM, and 22% for deep GM. One participant (submission 18, Figure 1) measured a large dataset (13 individuals) on three scanners and two vendors, and the intra-submission COVs for this submission were 3.2% for genu, 3.1% for splenium, 6.9% for cortical GM, and 7.1% for deep GM. The binomial appearance for the splenium, deep GM, and cortical GM for the sites used in the inter-site analyses (green) can be explained by an outlier measurement, which can be seen in (Figure 4 e-f, site 3.001).
Figure 3:Summary of results of the challenge as violin plots displaying the inter- and intra- submission dataset comparisons for phantoms (a) and human brains (b). Green indicates submissions used for inter-submission analyses, and orange indicates the sites used for intra-submission analyses.
A scatterplot of the T1 data for all submissions and their ROIs is shown in Figure 4 (phantom a-c, and human brains d-f). The NIST phantom T1 measurements are presented in each plot for different axes types (linear, log, and error) to better visualize the results. Figure 4-a shows good agreement for this dataset in comparison with the temperature-corrected reference T1 values. However, this trend did not persist for low T1 values (T1 < 100-200 ms), as seen in the log-log plot (Figure 4-b), which was expected because the imaging protocol is optimized for human brain T1 values (T1 > 500 ms). Higher variability is seen at long T1 values (T1 ~ 2000 ms) in Figure 4-a. Errors exceeding 10% are observed in the phantom spheres with T1 values below 300 ms (Figure 4-c), and 3-4 measurements with outlier values exceeding 10% error were observed in the human brain tissue range (~500-2000 ms).
Figure 4 d-f displays the scatter plot data for human datasets submitted to this challenge, showing mean and standard deviation T1 values for the WM (genu and splenium) and GM (cerebral cortex and deep GM) ROIs. Mean WM T1 values across all submissions were 828 ± 38 ms in the genu and 852 ± 49 ms in the splenium, and mean GM T1 values were 1548 ± 156 ms in the cortex and 1188 ± 133 ms in the deep GM, with less variations overall in WM compared to GM, possibly due to better ROI placement and less partial voluming in WM. The lower standard deviations for the ROIs of human database ID site 9 (by submission 18, Figure 1, and seen in orange in Figure 4d-g) are due to good slice positioning, cutting through the AC-PC line and the genu for proper ROI placement, particularly for the corpus callosum and deep GM.
Figure 6:Measured mean T1 values vs. temperature-corrected NIST reference values of the phantom spheres are presented as linear plots (a), log-log plots (b), and plots of the error relative to the reference T1 value (c). Green indicates submissions used for inter-submission analyses, and orange indicates the sites used for intra-submission analyses. The dashed lines in plots (c) represent a ±10 % error. Mean T1 values in two sets of ROIs, white matter (one 5⨯5 voxel ROI for genu, one 5x5 voxel ROI for splenium) and gray matter (three 3⨯3 voxel ROIs for cortex, one 5x5 voxel ROI for deep GM). In subplot g), the missing datapoints for deep GM for sites 1, 8, and 10 were due to the slice positioning of the acquisition not containing deep GM.
3.1| discussion¶
This challenge focused on exploring if different research groups could reproduce T1 maps based on the protocol information reported in a seminal publication Barral et al., 2010. Eighteen submissions independently implemented the inversion recovery T1 mapping acquisition protocol as outlined in Barral et al. Barral et al., 2010, and reported T1 mapping data in a standard quantitative MRI phantom and/or human brains at 27 MRI sites, using systems from three different vendors (GE, Philips, Siemens). The collaborative effort produced an open-source database of 94 T1 mapping datasets, including 38 ISMRM/NIST phantom and 56 human brain datasets. The inter-submission variability was twice as high as the intra-submission variability in both phantom and human brain T1 measurements, demonstrating that written instructions communicated via a PDF are not enough for reproducibility in quantitative MRI. This study reports the inherent uncertainty in T1 measures across independent research groups, which brings us one step closer to producing a practical baseline of variations for this metric.
Overall, our approach did show improvement in the reproducibility of T1 measurements in vivo compared to researchers implementing T1 mapping protocols completely independently (i.e. with no central guidance), as literature T1 values in vivo vary more than reported here (e.g., Bojorquez et al. Bojorquez et al., 2017 reports that reported T1 values in WM vary between 699 to 1735 ms in published literature). We were aware that coordination was essential for a quantitative MRI challenge, which is why the protocol specifications we provided to researchers were more detailed than any guidelines for quantitative MR imaging that were available at the time. Yet, even in combination with the same T1 mapping processing tools, this level of description (a PDF + post-processing tools) leaves something to be desired.
This analysis highlights that more information is needed to unify all the aspects of a pulse sequence across sites, beyond what is routinely reported in a scientific publication. However, in a vendor-specific setting, this is a major challenge given the disparities between proprietary development libraries Gracien et al., 2020. Vendor-neutral pulse sequence design platforms Layton et al., 2017Cordes et al., 2020Karakuzu et al., 2022 have emerged as a powerful solution to standardize sequence components at the implementation level (e.g., EF pulse shape, gradients, etc.). Vendor neutrality has been shown to significantly reduce the variability of T1 maps acquired using VFA across vendors Karakuzu et al., 2022. In the absence of a vendor-neutral framework, a vendor-specific alternative is the implementation of a strategy to control the saturation of MT across TRs A G Teixeira et al., 2020. Nevertheless, this approach can still benefit from a vendor-neutral protocol to enhance accessibility and unify implementations. This is because vendor-specific constraints are recognized to impose limitations on the adaptability of sequences, resulting in significant variability even when implementations are closely aligned within their respective vendor-specific development environments Lee et al., 2019.
The 2020 Reproducibility Challenge, jointly organized by the Reproducible Research and Quantitative MR ISMRM study groups, led to the creation of a large open database of standard quantitative MR phantom and human brain inversion recovery T1 maps. These maps were measured using independently implemented imaging protocols on MRI scanners from three different manufacturers. All collected data, processing pipeline code, computational environment files, and analysis scripts were shared with the goal of promoting reproducible research practices, and an interactive dashboard was developed to broaden the accessibility and engagement of the resulting datasets (https://
By providing access and analysis tools for this multi-center T1 mapping dataset, we aim to provide a benchmark for future T1 mapping approaches. We also hope that this dataset will inspire new acquisition, analysis, and standardization techniques that address non-physiological sources of variability in T1 mapping. This could lead to more robust and reproducible quantitative MRI and ultimately better patient care.
Acknowledgments¶
The conception of this collaborative reproducibility challenge originated from discussions with experts, including Paul Tofts, Joëlle Barral, and Ilana Leppert, who provided valuable insights. Additionally, Kathryn Keenan, Zydrunas Gimbutas, and Andrew Dienstfrey from NIST provided their code to generate the ROI template for the ISMRM/NIST phantom. Dylan Roskams-Edris and Gabriel Pelletier from the Tanenbaum Open Science Institute (TOSI) offered valuable insights and guidance related to data ethics and data sharing in the context of this international multi-center conference challenge. The 2020 RRSG study group committee members who launched the challenge, Martin Uecker, Florian Knoll, Nikola Stikov, Maria Eugenia Caligiuri, and Daniel Gallichan, as well as the 2020 qMRSG committee members, Kathryn Keenan, Diego Hernando, Xavier Golay, Annie Yuxin Zhang, and Jeff Gunter, also played an essential role in making this challenge possible. We would also like to thank the Canadian Open Neuroscience Platform (CONP), the Quebec Bioimaging Network (QBIN), and the Montreal Heart Institute Foundation for their support in creating the NeuroLibre preprint. Finally, we extend our thanks to all the volunteers and individuals who helped with the scanning at each imaging site.
The authors thank the ISMRM Reproducible Research Study Group for conducting a code review of the code (Version 1) supplied in the Data Availability Statement. The scope of the code review covered only the code’s ease of download, quality of documentation, and ability to run, but did not consider scientific accuracy or code efficiency.
Lastly, we acknowledge use of ChatGPT (v3), a generative language model, for accelerating manuscript preparation. The co-first authors employed ChatGPT in the initial draft for transforming bullet point sentences into paragraphs, proofreading for typos, and refining the academic tone. ChatGPT served exclusively as a writing aid, and was not used to create or interpret results.
+++
4Data availability statement¶
An interactive NeuroLibre preprint of this manuscript is available at Boudreau et al. (2023). All imaging data submitted to the challenge, dataset details, registered ROI maps, and processed T1 maps are hosted on OSF https://
5References¶
Submissions were reviewed by MB and AK. Submission guidelines (README.md) and a GitHub issue checklist (
.github /ISSUE _TEMPLATE /data -submission -request .md) were checked. Lastly, the submitted data was passed to the T1 processing pipeline and verified for quality and expected values. Feedback was sent to the authors if their submission did not adhere to the requested guidelines, or if issues with the submitted datasets were found and if possible, corrected (e.g., scaling issues between inversion time data points). Strictly speaking, not all manufacturers operate at 3.0 T. Even though this is the field strength advertised by the system manufacturers, there is some deviation in actual field strength between vendors. The actual center frequencies are typically reported in the DICOM files, and these were shared for most datasets and are available in our OSF.io repository (https://
osf .io /ywc9g/). From these datasets, the center frequencies imply participants that used GE and Philips scanners were at 3.0T ( ~127.7 MHz), whereas participants that used Siemens scanners were at 2.89T (~123.2 MHz). For simplicity, we will always refer to the field strength in this article as 3T.The T1 values vs temperature tables reported by the phantom manufacturer did not always exhibit a linear relationship. We explored the use of spline fitting on the original data and quadratic fitting on the log-log representation of the data, Both methods yielded good results, and we opted to use the latter in our analyses. The code is found here, and a Jupyter Notebook used in temperature interpolation development is here.
Only T1 maps measured using phantom version 1 were included in this inter-submission COV, as including both sets would have increased the COV due to the differences in reference T1 values. There were seven research groups that used version 1, and six that used version 2.
- Keenan, K. E., Biller, J. R., Delfino, J. G., Boss, M. A., Does, M. D., Evelhoch, J. L., Griswold, M. A., Gunter, J. L., Hinks, R. S., Hoffman, S. W., Kim, G., Lattanzi, R., Li, X., Marinelli, L., Metzger, G. J., Mukherjee, P., Nordstrom, R. J., Peskin, A. P., Perez, E., … Sullivan, D. C. (2019). Recommendations towards standards for quantitative MRI (qMRI) and outstanding needs. J. Magn. Reson. Imaging, 49(7), e26–e39.
- Fryback, D. G., & Thornbury, J. R. (1991). The efficacy of diagnostic imaging. Med. Decis. Making, 11(2), 88–94.
- Schweitzer, M. (2016). Stages of technical efficacy: Journal of Magnetic Resonance Imaging style. J. Magn. Reson. Imaging, 44(4), 781–782.
- Seiberlich, N., Gulani, V., Campbell, A., Sourbron, S., Doneva, M. I., Calamante, F., & Hu, H. H. (2020). Quantitative Magnetic Resonance Imaging. Academic Press.
- Damadian, R. (1971). Tumor detection by nuclear magnetic resonance. Science, 171(3976), 1151–1153.
- Pykett, I. L., & Mansfield, P. (1978). A line scan image study of a tumorous rat leg by NMR. Phys. Med. Biol., 23(5), 961–967.
- Stikov, N., Boudreau, M., Levesque, I. R., Tardif, C. L., Barral, J. K., & Pike, G. B. (2015). On the accuracy of T1 mapping: Searching for common ground. Magn. Reson. Med., 73(2), 514–522.
- Boudreau, M., Keenan, K. E., & Stikov, N. (2020). Quantitative T1 and T1r Mapping. In Quantitative Magnetic Resonance Imaging (pp. 19–45).
- Bottomley, P. A., Foster, T. H., Argersinger, R. E., & Pfeifer, L. M. (1984). A review of normal tissue hydrogen NMR relaxation times and relaxation mechanisms from 1-100 MHz: dependence on tissue type, NMR frequency, temperature, species, excision, and age. Med. Phys., 11(4), 425–448.
- Wansapura, J. P., Holland, S. K., Dunn, R. S., & Ball, W. S., Jr. (1999). NMR relaxation times in the human brain at 3.0 tesla. J. Magn. Reson. Imaging, 9(4), 531–538.
- Dieringer, M. A., Deimling, M., Santoro, D., Wuerfel, J., Madai, V. I., Sobesky, J., von Knobelsdorff-Brenkenhoff, F., Schulz-Menger, J., & Niendorf, T. (2014). Rapid parametric mapping of the longitudinal relaxation time T1 using two-dimensional variable flip angle magnetic resonance imaging at 1.5 Tesla, 3 Tesla, and 7 Tesla. PLoS One, 9(3), e91318.
- Ernst, R. R., & Anderson, W. A. (1966). Application of Fourier Transform Spectroscopy to Magnetic Resonance. Rev. Sci. Instrum., 37(1), 93–102.
- Redpath, T. W., & Smith, F. W. (1994). Technical note: use of a double inversion recovery pulse sequence to image selectively grey or white brain matter. Br. J. Radiol., 67(804), 1258–1263.
- Tofts, P. S. (1997). Modeling tracer kinetics in dynamic Gd-DTPA MR imaging. J. Magn. Reson. Imaging, 7(1), 91–101.
- Sled, J. G., & Pike, G. B. (2001). Quantitative imaging of magnetization transfer exchange and relaxation properties in vivo using MRI. Magn. Reson. Med., 46(5), 923–931.